Software and algorithmic complex for cuffless arterial pressure monitoring based on synchronous recording of local pulse wave velocity and arterial diameter changes
Type of article: scientific article
DOI: https://doi.org/10.18127/j15604136-202606-18
UDC: 615.47:616-072.7
Authors:

A.A. Khomyakova1, M.A. Yangirov2, M. Al-harosh3

1,2,3 FSAEI HE “Bauman Moscow State Technical University” (Moscow, Russia)

1 khomyakovaaa@student.bmstu.ru

Abstract:

This paper presents an algorithmic and software complex for cuffless beat-to-beat arterial pressure monitoring using synchronous two-channel electrical impedance plethysmography to record local pulse wave velocity and arterial diameter changes. Analytically, the artery is modelled as a homogeneous cylindrical inclusion that maps the arterial radius to impedance, while the pressure–diameter relationship follows the haemodynamic P–β exponential model. Data acquisition uses a tetrapolar front-end sampling at 500 Hz, processed by a four-stage Python workflow: robust Akima-spline detrending with automatic artefact rejection; a six-point fiducial detector on up-sampled data that determines beat-to-beat pulse wave velocity; an accelerated Brent-bracketed solver for the inverse impedance problem; and a median-anchored two-point pressure calibration. The complex was verified using in vivo recordings from a healthy volunteer. After artefact removal, the reconstructed beat-to-beat pressures varied around the calibration values of 122 mmHg (systolic) and 75 mmHg (diastolic), and the average runtime per beat did not exceed 0.7 s. The results expose the physiological cycle-to-cycle variability that conventional cuff-based devices inherently average out, and provide the algorithmic basis of a wearable cuffless blood pressure monitoring system.

Pages: 106-114
For citation

Khomyakova A.A., Yangirov M.A., Al-harosh M. Software and algorithmic complex for cuffless arterial pressure monitoring based on synchronous recording of local pulse wave velocity and arterial diameter changes // Biomedical Radioelectronics. 2026. V. 29. № 6. P. 106−114. DOI: https://doi.org/10.18127/j15604136-202606-18

References
  1. Oliver J.J., Webb D.J. Noninvasive assessment of arterial stiffness and risk of atherosclerotic events. Arterioscler Thromb Vasc Biol. Apr. 2003. V. 23. № 4. P. 554–566. doi: 10.1161/01.ATV.0000060460.52916.D6.
  2. McEniery C. M., Cockcroft J.R., Roman M.J., Franklin S.S., Wilkinson I.B. Central blood pressure: current evidence and clinical importance. European Heart Journal. 2014. V. 35. № 26. P. 1719–1725.
  3. Ding X.-R. et al. Continuous blood pressure measurement from invasive to unobtrusive: celebration of 200th birth anniversary of Carl Ludwig. IEEE J Biomed Health Inform, Nov. 2016. V. 20. № 6. P. 1455–1465.
  4. Sharma M. et al. Cuff-less and continuous blood pressure monitoring: a methodological review. Technologies. May 2017. V. 5. № 2. P. 1–22.
  5. Mukkamala R. et al. Toward ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans Biomed Eng. Aug. 2015. V. 62. № 8. P. 1879–1901.
  6. Ding X. et al. Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation. Physiol Meas. Feb. 2016. V. 37. № 2. P. 227–237.
  7. Nabeel P.M., Joseph J., Karthik S., Sivaprakasam M., Chenniappan M. Bi-modal arterial compliance probe for calibration-free cuffless blood pressure estimation. IEEE Trans Biomed Eng, Nov. 2018. V. 65. № 11. P. 2392–2404.
  8. Al-harosh M., Yangirov M., Kolesnikov D., Shchukin S. Bio-impedance sensor for real-time artery diameter waveform assessment. Sensors. 2021. V. 21. № 24. Art. № 8438. doi: 10.3390/s21248438.
  9. Sugawara M., Niki K., Furuhata H., Ohnishi S., Suzuki S. Relationship between the pressure and diameter of the carotid artery in humans. Heart Vessels. 2000. V. 15. P. 49–51.
  10. Smirnova A., Belsheva M., Mugeb A. Applicability assessment of electrical impedance plethysmography for vascular diseases diagnosis. In Proc. 2023 Systems and Technologies of the Digital HealthCare (STDH), IEEE. 2023. P. 118–122.
  11. Al-haroshM., Chernikov E., Shchukin S. Patient-specific numerical modeling for renal blood monitoring using electrical bio-impedance. Sensors. 2022. V. 22. № 2. Art. no. 606. DOI: 10.3390/s22020606.
  12. Bramwell J.C., Hill A.V. The velocity of the pulse wave in man. Proc. Roy. Soc. Lond. B. 1922. V. 93. № 652. P. 298–306.
Date of receipt: 03.08.2026
Approved after review: 17.08.2026
Accepted for publication: 28.08.2026