350 rub
Journal Electromagnetic Waves and Electronic Systems №7 for 2014 г.
Article in number:
The sensitivity of a photoreceiver based on a photocurrent integrator
Authors:
V.D. Kuptsov - Ph. D. (Eng.), St. Petersburg State Polytechnical University. E-mail: vdkuptsov@yandex.ru
V.P. Valyukhov - Dr. Sc. (Eng.), Associate Professor, St. Petersburg State Polytechnical University. E-mail: valyukhov@yandex.ru
Abstract:
The problem of calculating the threshold sensitivity of an integrating photoreceiver (photocurrent integrator) is investigated. Photocurrent integrators are used in scientific instruments to achieve high sensitivity of an optical measurement system in case of relatively slow physical processes. Noise model of a photointegrator is used to develop equivalent noise circuits and to calculate the RMS value of the noise voltage across the photoreceiver's output. Three types of approximations are investigated: a real photointegrator; a photointegrator with an ideal integrating capacitor and a reset switch; an ideal photointegrator. The threshold sensitivity of a photointegrator is defined as the power at the input of a photointegrator at which the root mean squared (RMS) voltage across its output is equal to the RMS voltage of total noise i.e. the signal to noise ratio is equal to one. A photomultiplier tube (PMT) can be used to increase the sensitivity. Formulas for calculating the sensitivity of a PMT based photointegrator are given. The increase in sensitivity due to the use of a PMT can be up to 18−30 times. Experimental study of a photometer of a molecular condensation nuclei (MCN) detector that forms a base of highly sensitive MCN gas analyzers was conducted. Measured sensitivity differed from calculated by 10%, however at femtowatt power levels it is very difficult to get rid of parasitic optical signals that are responsible for a small decrease in sensitivity compared to the theoretical prediction. It is noted that in many practical applications, for example in X ray absorption method of mineral extraction using position sensitive photodiode X ray receivers, an approximation of a real photointegrator must be used.
Pages: 16-23
References

  1. Chernov E.I.  Precision photoreceivers based on photodiodes. Part II // Reviews of consumer electronics. Ser. 4: Vacuum and gas-discharge devices. 1989.  №1420. 52 P.
  2. Zaezdny А.МFundamentals of statistical radio engineering.  M: Svyaz. 1969.  448 P.
  3. Kuptsov V.D., Valyukhov V.P. Sensitivity of fiber optic communication lines photoreceivers//Scientific and technological information of the SPbSPU. 2010.V.113.№ 6.  P.31-37.
  4. Kuptsov V.D., Katelevsky V.Ya., Valyukhov V.P. Optoelectronic devices for molecular condensation nuclei effect gas analyzers. SPbSPU Publishing. 2012.142 P.
  5. Kuptsov V.D., Katelevsky V.Ya., Valyukhov V.P. Optoelectronic devices for molecular condensation nuclei effect gas analyzers. Electronic resource. Monograph. SPbSPU. 2012. URL: http://www.unilib.neva.ru/dl/2618.pdf.
  6. Kyandzhetsian R.A., Katelevsky V.Ya., Valyukhov V.P., Kuptsov V.D., Demin S.V.Highly sensitive gas analyzers based on molecular condensation nuclei detector for ecological monitoring and testing of highly effective filters//Int. J. Alternative Energy Ecol.2011.V.102. №10. P.32-42.
  7. Kuptsov V.D., Kyandzhetsian R.A., Katelevsky V.Ya., Valyukhov V.P. Gas analyzers based on the  molecular condensation nuclei effect//Scientific and technological information of the SPbSPU.2010.V.113.  № 6.  P.145-151.
  8. Rybin E.N., Valyukhov V.P., Kuptsov V.D. Thermodynamics of Nucleation of Supersaturated Vapor on Molecular Condensation Centers // Technical Physics. The Russian Journal of Applied Physics. 2012. V.57. №8.  P.1062−1067.
  9. Rybin E.N., Valyukhov V.P., Kuptsov V.D. Simulation of Supersaturated Vapor Nucleation on Molecular Condensation Centers // Technical Physics.The Russian Journal of Applied Physics. 2012. V.57. № 8. P.1068−1074.
  10. Kuptsov V.D.,  Kyandzhetsian  R.A.,  Katelevsky  V.Ya., Valyukhov V.P.Light scattering by aerosol particles in gas analyzers based on molecular condensation nuclei//Scientific and technological information of the SPbSPU.2011.  V.115.  № 1.  P.178-187.
  11. Kuptsov V.D., Katelevskii V.Y., Valyukhov V.P.,Rybin E.N. Optoelectronic system of the aerosol photometer in the detector of molecular condensation nuclei // Technical Physics. The Russian Journal of Applied Physics. 2013. V. 58.  № 8.  P.1211−1218.
  12. Kuptsov V.D., Katelevskii V.Y., Valyukhov V.P.,Rybin E.N. Optoelectronic system of the aerosol photometer in the detector of molecular condensation nuclei // JTF. 2013. V. 83. № 8.  P.127−134.
  13. Demin S.V., Kuptsov V.D., Valyukhov V.P., Katelevsky V.Ya. Gas analyzers based on molecular condensation nuclei for detection of poisonous substances// Scientific and technological information of the SPbSPU.  2011.V.120.  №.2.  P.183-191.
  14. Patent  RU 2 472 595: The method of separation of minerals / Novikov V.V., Rudakov V.V., Zlobin M.N., Malakhovsky V.I., Kuptsov V.D., Valyukhov V.P., Kuptsov M.V., Olchovoj V.A.; Declared 14.12.2011, published 20.01.2013. Bulletin of inventions №2. 11 P.