350 rub
Journal Biomedical Radioelectronics №4 for 2018 г.
Article in number:
Measuring T2 relaxation time in MRI gradient echo sequences
Type of article: scientific article
UDC: 537.635
Authors:

In magnetic resonance imaging (MRI), the gradient recalled echo (GRE) sequence became a powerful tool for classification of biological tissues due its sensitivity to structural heterogeneities of mesoscopic and macroscopic scale. The atomic scale heterogeneities, determined by spin-spin interaction of protons, can be imaged in contrast of transverse relaxation time T2, obtained in spin-echo (SE) sequence. For many years, the SE sequence was the only way to measure T2 accurately because, in GRE sequence, this parameter is masked by more rapid decay of the signal due to magnetic gradients. First attempts to evaluate - even not exactly measure - T2 in GRE sequences revealed serious algorithmic problems. Recently introduced new solution to this problem - multipoint algorithm with clamping (MPC) - opens the possibility for accurate and fast measurements of T2 in GRE sequences. The present short communication presents quantitative statistical confirmation of its performance on phantoms. It is shown that MPC algorithm produces accurate measurements of T2 in GRE sequences with average discrepancies of about 2% on small magnetic heterogeneities and 8% on the large ones with Pearson coefficients in the range 0.9-0.95. Commonly accepted in practice evaluation of T2 on GRE sequences by means of monoexponential fitting (the so-called  values) gives inadequate average values with errors of about 30% and Pearson coefficient of only 0.7.

Abstract:

In this article, a possibility to measure separately both magnetic gradients and transverse relaxation time T2 in gradient-recalled echo (GRE) sequence only, without application of spin-echo (SE) sequence was considered. Statistical analysis of phantom images in a form of two-dimensional graph together with single-value Pearson coefficient present quantitative confirmation for correctness of measuring T2 in GRE sequences.

Pages: 33-37
References
  1. Edelstein W.A., Hutchison J.M., Johnson G., Redpath T. Spin warp NMR imaging and applications to human whole-body imaging // Phys. Med. Biol. 1980. V.25. P. 751–756.
  2. Haase A., Matthaei D., Hanicke W., Frahm J. Dynamic digital subtraction imaging using fast low-angle shot MR movie sequence // Radiology. 1986.
    V. 160. P. 537–541.
  3. Hahn E.L. Spin echoes // Physical Review. 1950. V. 80(4). P. 580–594.
  4. Fernandez-Seara M.A., Wehrli F.W. Postprocessing tech-nique to correct for background gradients in image-based R2* measurements // Magn. Reson. Med. 2000. V. 44. P. 358–366.
  5. Dahnke H., Schaeffter T. Limits of detection of SPIO at 3.0T using T2* relaxometry // Magn. Reson. Med. 2005. V. 53. P. 1202–1206.
  6. Protopopov A. Izmerenie gradientov magnitnogo polya v MRT. Ch. 1. Model i algoritmy // Elektromagnitnye volny i elektronnye sistemy. 2017. T.22(1). S. 37–44.
  7. Protopopov A. Izmerenie gradientov magnitnogo polya v MRT. Ch. 2. Eksperimenty na fantomah // Elektromag-nitnye volny i elektronnye sistemy. 2017. T. 22(2). S. 62–67.
  8. Reichenbach J.R., Hacklander Th., Harth T., Hofer M., Rassek M., Modder U. 1H T1 and T2 measurements of the MR imaging contrast agents Gd-DTPA and Gd-DTPA BMA at 1.5T // Eur. Radiol. 1997. V.7. P. 264–274.
  9. Rohrer M. Comparison of magnetic properties of MRI con-trast media solutions at different magnetic field strengths // Investigative Radiology. 2005. V. 40(11). P. 715–724.
Date of receipt: 4 марта 2018 г.