350 rub
Journal Biomedical Radioelectronics №7 for 2010 г.
Article in number:
Digital Electrostatical Fluxmeter
Authors:
V.A. Efimov
Abstract:
A fluxmeter has been developed, which is different from known ones in the thing that some measures of error re-duction have been realized. Using digital method of measurement of signal amplitude of measured current from input transformer allows to exclude errors connected whith the instrument drift of the operational amplifieres and also to solve the problem offorming digitals of the experimental data and transfering them to a receiving computer. Such devices can be united in a system. The fluxmeter allows to measure electric intensity within 10000 v/m with an averege square value of measuring error 1v/m. The service band is 5 Hz, supply voltage is 12V.
Pages: 58-64
References
  1. Чалмерс Дж.А.Атмосферное электричество / пер. с англ. под ред. И.М. Имянитова. Л.: Гидрометеоиздат. 1974. 420 с.
  2. Имянитов И.М.Приборы и методы для изучения электричества атмосферы. М.: ГТТИ. 1957. 483 с.
  3. Мусин Р.Ф., Морозов В.А., Матлашов А.Н. Модуляционный измеритель напряженности слабых квазистатических электрических полей // Приборы и техника эксперимента. 1985. № 4. С.127 - 128.
  4. Гер А.А., Левин А.С., Носов Ю.Р. Электрометрический варикап. / Полупроводниковые приборы и их применение. Вып. 28. М.: Сов. радио. 1974.
  5. Радиотехнические системы: Учеб. для вузов / под ред. Ю.М.Казаринова. М.: Высш. шк. 1990. 496 с.