E.N. Rubtsova1, A.A. Pugovkin2, A.V. Koroteev3, D.V. Telyshev4
1,4 Institute of Biomedical Systems, National Research University of Electronic Technology (Moscow, Russia)
1–4 I.M. Sechenov First Moscow State Medical University (Moscow, Russia)
1 rubtsova_e_n@staff.sechenov.ru, 2 pugovkin_a_a@staff.sechenov.ru, 3 koroteev_a_v@staff.sechenov.ru,
4 telyshev_d_v@staff.sechenov.ru
Univentricular palliation is performed in patients with congenital heart defects, such as single functional ventricle, hypoplasia, septal defects, valve defects, obstructive defects, etc. After the palliation, the single ventricle becomes systemic, and venae cavae are directly or indirectly connected to pulmonary arteries. Palliated patients may temporary lead a full life yet eventually need heart transplantation. Mechanical circulatory support (MCS) may be an alternative treatment for heart transplantation in patients having univentricular circulation. Personalized modeling of a biotechnical MCS system allows to forecast current and optimal system states and assess support advisability and efficacy before implanting a circulatory assist device. In this study, a mathematical model of univentricular circulation with possibility to connect a circulatory assist device is presented. An algorithm for efficacy forecast and assessment of personalized mechanical univenricular circulatory support was developed. The algorithm automatically determines circulatory model parameters based on patient’s hemodynamic data and a set of parameter templates for corresponding age groups and circulatory states. This first stage is based on genetic algorithm with model parameters used as a genotype. Criterion of obtaining personalized model parameters is a zero sum of relative deviations of patient’s hemodynamic data and simulated data after personalization given the acceptable ranges of deviations determined based on accuracy of clinical diagnostic devices. At the second stage, the developed algorithm simulates a personalized biotechnical system of mechanical univentricular circulatory support using personalized circulatory model and pressure head–flow rate characteristics of circulatory assist device. At the third stage, the algorithm assesses advisability of chosen control strategy with negative recommendation for inadvisable strategy. At the last stage, the algorithm forecasts optimal states of biotechnical system determined as sufficiently approximate to normal circulation. Verification of the developed algorithm was performed with patients’ hemodynamic data from literature. Distributions of hemodynamic parameters in simulated biotechnical systems for three patients are described along with forecast optimal states and corresponding values of control parameter for chosen methods of support. In these cases, pediatric rotary blood pump previously developed by our research group was used as circulatory assist device. Assessment of changes in biotechnical system after implantation of circulatory assist device revealed inefficacy of MCS in one of three considered patients.
Rubtsova E.N., Pugovkin A.A., Koroteev A.V., Telyshev D.V. Efficacy assessment of strategies for mechanical univentricular circulatory support. Dynamics of complex systems. 2024. V. 18. № 1. P. 41−50. DOI: 10.18127/j19997493-202401-04 (in Russian).
- Lacour-Gayet F.G., Lanning C.J., Stoica S., Wang R., Rech B.A., Goldberg S., Shandas R. An Artificial Right Ventricle for Failing Fontan: In Vitro and Computational Study. The Annals of Thoracic Surgery. 2009. V. 88. № 1. P. 170–176.
- Petukhov D.S., Telyshev D.V. A Mathematical Model of the Cardiovascular System of Pediatric Patients with Congenital Heart Defect. Biomedical Engineering. 2016. V. 50. № 4. P. 229–232.
- Granegger M., Schweiger M., Daners M.S., Meboldt M., Hübler M. Cavopulmonary mechanical circulatory support in Fontan patients and the need for physiologic control: A computational study with a closed-loop exercise model. The International Journal of Artificial Organs. 2018. V. 41. № 5. P. 261–268.
- Farahmand M., Kavarana M.N, Kung E.O. Risks and Benefits of Using a Commercially Available Ventricular Assist Device for Failing Fontan Cavopulmonary Support: A Modeling Investigation. IEEE Trans Biomed Eng. 2020. V. 67. № 1. P. 213–219.
- Pekkan K., Frakes D., De Zelicourt D., Lucas C.W., Parks W.J., Yoganathan A.P. Coupling Pediatric Ventricle Assist Devices to the Fontan Circulation: Simulations with a Lumped-Parameter Model. ASAIO Journal. 2005. V. 51. № 5. P. 618–628.
- Rodefeld M.D., Frankel S.H., Giridharan G.A. Cavopulmonary assist: (Em) powering the univentricular Fontan circulation. Pediatric Cardiac Surgery Annual. 2011. V. 14. P. 45–54.
- Kung E., Pennati G., Migliavacca F., Hsia T.-Y., Figliola R., Marsden A., Giardini A. A Simulation Protocol for Exercise Physiology in Fontan Patients Using a Closed Loop Lumped-Parameter Model. J. Biomech Eng. 2014. V.136. № 8. P. 0810071–08100714.
- Kiselev I.N., Semisalov B.V., Biberdorf E.A., SHaripov R.N., Blohin A.M., Kolpakov F.A. Modul'noe modelirovanie serdechno-sosudistoj sistemy cheloveka. Matematicheskaya biologiya i bioinformatika. 2012. T. 7. № 2. S. 703–736.
- Capoccia M. Development and Characterization of the Arterial Windkessel and Its Role During Left Ventricular Assist Device Assistance. Artificial Organs. 2015. V. 39. № 8. P. 138–153.
- Segers P., Rietzschel E.R., De Buyzere M.L., Stergiopulos N., Westerhof N., Van Bortel L.M., Gillebert T., Verdonck P.R. Three- and four-element Windkessel models: assessment of their fitting performance in a large cohort of healthy middle-aged individuals. Proceedings of the Institution of Mechanical Engineers. 2008. V. 222. № 4 P. 417–428.
- Shi Y., Lawford P., Hose R. Review of Zero-D and 1-D Models of Blood Flow in the Cardiovascular System. BioMedical Engineering OnLine. 2011. V. 33. № 10. P. 1–38.
- Kokalari I., Karaja T., Guerrisi M. Review on lumped parameter method for modeling the blood flow in systemic arteries. Biomedical Science and Engineering. 2013. V. 6. № 1. P. 92–99.
- Faragallah G., Simaan M.A. An Engineering Analysis of the Aortic Valve Dynamics in Patients with Rotary Left Ventricular Assist Devices. Journal of Healthcare Engineering. 2013. V. 4. № 3. P. 307–327.
- Suga H., Sagawa K. Instantaneous pressure-Vume relationships and their ratio in the excised, supported canine left ventricle. Cirtulation Rararth. 1974. V. 35. № 1. P. 117–126.
- Pugovkin A.A., Markov A.G., Selishchev S.V., Korn L., Walter M., Leonhardt S., Bockeria L.A., Bockeria O.L., Telyshev D.V. Advances in Hemodynamic Analysis in Cardiovascular Diseases Investigation of Energetic Characteristics of Adult and Pediatric Sputnik Left Ventricular Assist Devices during Mock Circulation Support. Cardiol Res Pract. 2019. Article ID 4593174. P. 1–15.
- Telyshev D., Denisov M., Pugovkin A., Selishchev S., Nesterenko I. The Progress in the Novel Pediatric Rotary Blood Pump Sputnik Development. Artificial Organs. 2018. V. 42. № 4. P. 1–12.
- Ohuchi H., Miyazaki A., Negishi J., Hayama Y., Nakai M., Nishimura K. et al. Hemodynamic determinants of mortality after Fontan operation. Am Heart J. 2017. V. 189. P. 9–18.
- Di Molfetta A., Ferrari G., Filippelli S., Fresiello L., Iacobelli R., Gagliardi M.G., Amodeo A. Use of Ventricular Assist Device in Univentricular Physiology: The Role of Lumped Parameter Models. Artificial Organs. 2016. V. 40. № 5. P. 444–453.
- Barber G., Di Sessa T., Child J.S., Perloff J.K., Laks H., George B.L. et al. Hemodynamic responses to isolated increments in heart rate by atrial pacing after a Fontan procedure. Am Heart J. 1988. V. 115. № 4. P. 837–841.
- Hsu P.L., Wang D., Ballard-Croft C., Xiao D., Zwischenberger J.B. A numerical simulation comparing a cavopulmonary assist device and VA ECMO for failing Fontan support. ASAIO J. 2017. V. 63. P. 604–612.
- Voors A.W., Foster T.A., Frerichs R.R., Webber L.S., Berenson G.S. Studies of blood pressures in children, ages 5–14 years, in a total biracial community. The Bogalusa Heart Study. Circulation. 1976. V. 54. № 2. P. 319–327.
- Cheng C.P., Herfkens R.J., Taylor C.A., Feinstein J.A. Proximal pulmonary artery blood flow characteristics in healthy subjects measured in an upright posture using MRI: the effects of exercise and age. J. Magn. Res Imag. 2005. V. 21. P. 752–758.
- Cattermole G.N., Leung P.Y.M., Mak P.S.K., Chan S.S.W., Graham C.A., Rainer T.H. The normal ranges of cardiovascular parameters in children measured using the ultrasonic cardiac output monitor. Crit. Care Med. 2010. V. 38. № 9. P. 1875–1881.
- Levy P.T., Patel M.D., Groh G., Choudhry S., Murphy J., Holland M.R. et al. Pulmonary arterial acceleration time provides a reliable estimate of invasive pulmonary hemodynamics in children. J. Am. Soc. Echocardiogr. 2016. V. 29. № 11. P. 1056–1065.
- Cattermole G.N., Leung P.Y.M., Ho G.Y.L., Lau P.W.S., Chan C.P.Y., Chan S.S.W. et al. The normal ranges of cardiovascular parameters measured using the ultrasonic cardiac output monitor. Physiol Rep. 2017. V. 5. № 6. P. e13195.
- Hjortdal V.E., Christensen T.D., Larsen S.H., Emmertsen K., Pedersen E.M. Caval blood flow during supine exercise in normal and Fontan patients. Ann Thorac Surg. 2008. V. 85. P. 599–603.
- Porfiryev A., Markov A., Galyastov A., Denisov M., Burdukova O., Gerasimenko A.Yu., Telyshev D. Fontan Hemodynamics Investigation via Modeling and Experimental Characterization of Idealized Pediatric Total Cavopulmonary Connection. Appl. Sci. 2020. V. 10. Art. num. 6910. P. 1–17.